The pre-amplifier circuit and the rear-stage AD of the thermocouple circuit I designed have problems. How to modify the circuit?

Hardware design
10월 07, 2020 by Delon 1300

Add the thermocouple signal directly to the rear electrode AD. It is clearly a positive voltage mv signal. The measured AD is indeed negative. FFF6 fluctuates upwards. There is also a problem with the pre-amplification. Touch the hot end of the thermocouple with a voltmeter. The measured output signal keeps rising, but it really drops after being amplified by the meter. I don’t know what to do? 18b20 for the cold end.

모든 댓글

user image

Misa 게시 날짜 October 7, 2020

If there is a signal source, you can use the signal source to simulate the sensor input on the input terminal. If the sensor is actually connected, you may want to see how your shield is connected.

0
user image

Alphonse 게시 날짜 October 7, 2020

You can debug the two levels separately, adjust AD first,

Is your AD acquisition voltage OK? You can input a voltage to see the output, right?

There should be no problem up front, an instrumentation amplifier.

0
user image

Serra 게시 날짜 October 7, 2020

Don’t trust the data after AD too much,

Directly use a high-precision voltmeter or oscilloscope to measure the voltage and waveform across C17 and the thermocouple,

Also pay attention to the input common mode voltage of AD620, because it is single-ended power supply. The REF pin of AD620 cannot change the common-mode voltage range of the AD620 input terminal, but can only change the common-mode voltage of the output signal.

0